Method for minimizing the errors of a capacitive self-compensating moisture meter
Abstract
The accuracy of measurements of a capacitive moisture-measuring system essentially depends on the extent to which it is possible with its help to measure only the reactive component of the total conductivity of the sample being monitored. An effective method for reducing the error component of the moisture-measuring system (MMS) caused by the influence of active losses and other uncontrolled factors is to measure them and, on the basis of the data obtained, introduce a correction into the output signal of the system.
Область исследования:
(Archived) Without section
Funding:
None
References
- Kotchenko F. F. Tracking Systems of Automatic Compensators. L., Nedra, 1965 (in Russian)
- Krichevsky E. S. High-Frequency Control of Moisture Content in Mineral Processing. M., Nedra, 1972 (in Russian)