Obtaining the calibration signal during analytical works on NMR-spectrometer JNM-3H-60, VN-2
Authors:
Abstract
The high sensitivity of the NMR spectrometer JNM-3H, BH-2 (Japan electron optics laboratory Co. Ltd.) allows its use for express quantitative analysis of solids. One of the significant difficulties in this case is related to the recording of the calibration signal, which allows taking into account the analysis errors that arise as a result of uncontrolled changes in the hardware parameters of the spectrometer. First of all, these include changes in the goodness of the sensor circuit due to the variability of the electrical properties of the substance under study and changes in the gain of the receiver.
Область исследования:
(Archived) Without section
Similar articles
Excavator characteristics of asynchronous electric drive with non-contact parametric regulation
1968 Yu. I. Fel
Basic parameters of the coal combine with adjustable and non-adjustable drive at destruction of coal of variable strength
1968 S. A. Shteding
Selection of the structure of the automatic control system (ACS) of the phosphorite concentrate drying process
1968 S. V. Storozhenko