Stabilization of the energy scale of the spectrometer using a gamma reference during testing of radioactive ores
Abstract
The electronic energy scale stabilization circuit is a device that monitors the photopeak of a gamma ray repeater. The basis of the scheme are two adjacent differential channels with the same window width, tuned, respectively, to the left and right branches of the photopeak of the reference source. At the output of both differential channels is placed a comparison circuit and an actuator...
Область исследования:
(Archived) Without section
References
- Matveev V. V., Khazanov B. I. Instruments for measuring ionizing radiation. Atomizdat, 1967.
- Novikov G. F., Sinitsyn A. Ya. Experience of using LSU-5k analyzer for determination of U(Ra), Th and K in boreholes. Geophysical Apparatus, 1968, issue 35.
- Novikov G. F., Sinitsyn A. Ya., Kozynda Yu. O. Optimal energy windows of scintillation gamma-spectrometer. Journal of Mining Institute, 1969, vol. LVI, issue 2.
- Novikov G. F., Sinitsyn A. Y., Kozynda Yu. O. Scintillation spectra of gamma-radiation of radioactive ores in different size crystals. NaJ (T1). Questions of ore geophysics, 1968, issue 8.
Similar articles
Continuous recording of equivalent uranium and thorium contents in gamma-spectral logging
1974 Yu. O. Kozynda, G. F. Novikov, A. Ya. Sinitsyn
Carbonate method of prospecting for Iceland spar deposits
1974 A. V. Skropyshev, L. M. Partashnikov, Zh. A. Nikolina
Angular distribution of scattered gamma-radiation of Cs137 source at density gamma-gamma method
1974 E. P. Popov, E. Kh. Vishnyakov
Consideration of borehole conditions in interpretation of gamma-spectral logs
1974 G. F. Novikov, Yu. O. Kozynda, A. Ya. Sinitsyn, B. V. Kalinin